Gauging performance of biosynthesized silver nanoparticles loaded polypropylene nonwoven based textile electrodes for 3-lead health monitoring electro cardiogram on analogous system
Source abstract
Abstract
This work reports the engineering of textile electrodes, considered safe for humans even if worn next to skin for a longer time. Obliging this phenomenon conductive Silver nanoparticles (AgNPs) were biosynthesized from Silver Sulphate (Ag 2 SO 4 ) and medicinal values enriched Ocimum Sanctum (Tulsi)leaves extract. These conductive Silver nanoparticles were loaded by spray technique on polypropylene nonwoven fabric having inbuilt antifungal characteristics, to reduce its resistivity (10Ω) for the fabrication of textile electrodes. The adequate skin-electrode impedance values were observed for the fabricated textile electrodes, viz; 1.44 MΩ–1.83 MΩ and 1.01 MΩ–1.18 MΩ, in the dry and wet state respectively. The 3-lead health monitoring electrocardiograms (ECG) were obtained on the Analogous system with the textile electrodes; dry and wet state as well as gel electrodes. The cardiograms were also taken at a smaller triangle than usual, only for the high resistance textile electrodes. The wet electrodes have executed considerably better clarity of PQR wavelets than reference gel electrodes ECG plots, and their performance was found consistent when tested after six months’ time leap. However, higher motion artifacts caused in the case of dry electrodes have resulted in distorted PQR wavelets and the tracing became worsen with increased testing time leap. This was mainly due to the encapsulation of conductive AgNPs in the air voids of the fabric, increased resistivity. The cardiogram quality has not shown peculiar benefit for a higher heart pumping pressure at the smaller triangle in either of the cases.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced Sensor and Energy Harvesting Materials · Conducting polymers and applications · Neuroscience and Neural Engineering
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Textile Engineering, The Maharaja Sayajirao University of Baroda, Faculty of Technology & Engineering, The Maharaja Sayajirao University of Baroda, Vadodara, India
Department of Textile Chemistry, The Maharaja Sayajirao University of Baroda, Faculty of Technology & Engineering, The Maharaja Sayajirao University of Baroda, Vadodara, India
References (34)
Show source-provided references
1. 10.1080/0040516080259747910.1080/00405160802597479 ↗
2. Muthu Kumar N · J Techn Text · 2013 · 81
3. Patent US5178143A-Electrically conductive gel composition, 12 January 1993. www.google.com/patentUS5178143 (accessed 2 June 2021).
4. 10.1088/1361-665X/abc3fa10.1088/1361-665X/abc3fa ↗
5. Hoffmann KP · In: Annu Int Conf IEEE Eng Med Biol Soc · 2007 · 1 · 5740
6. 10.1016/j.compositesa.2019.10568510.1016/j.compositesa.2019.105685 ↗
7. 10.1016/j.cej.2021.12868110.1016/j.cej.2021.128681 ↗
8. 10.1177/152808371243806910.1177/1528083712438069 ↗
9. 10.1039/c3tb00448a10.1039/c3tb00448a ↗
10. Saritha C · Bul. J Phys · 2008 · 35 · 68
11. Taniguchi N. · 1974
12. 10.1002/0470020873.ch110.1002/0470020873.ch1 ↗
13. 10.1021/acsami.0c1640110.1021/acsami.0c16401 ↗
14. Pereira T, Carvalho H, et al. Wearable biopotential measurement using the TIADS1198 analog front-end and textile electrodes. 2013 IEEE International Symposium on Medical Measurements and Applications (MeMeA) 2013; 325–330. Doi. 10.1109/MeMeA.2013.654976110.1109/MeMeA.2013.6549761 ↗
15. 10.1166/jgst.2015.103910.1166/jgst.2015.1039 ↗
16. Patel BH · Textile Asia · 2015 · 46 · 3
17. Márquez Ruiz JC. · 2011
18. 10.1002/app.3532310.1002/app.35323 ↗
19. Marco DR, Member IEEE, Francesco R, et al. MagIC system: a new textile-Based wearable device for biological signal monitoring. In: Applicability in daily life and clinical setting, engineering in medicine and biology 27th annual conference. Shanghai, China, 1–4 September 2005.
20. Bahareh T, Shervin S, Voicu G, et al. Impact of skin–electrode interface on electrocardiogram measurements using conductive textile electrodes. IEEE Trans Instrum Meas 2014; 63(6): 1412–1422.10.1109/TIM.2013.2289072 ↗
21. Ruffini G, Dunne S, Farres E, et al. A dry electrophysiology electrode using CNT arrays. Sens. Actuators A 2006; 132: 34–41. http://arxiv.org/abs/physics/0510110.1016/j.sna.2006.06.013 ↗
22. Ruffini G · Annu Int Conf IEEE Eng Med Biol Soc · 2007 · 1 · 6690
23. Ryu CY · 27th Annu Int Conf IEEE Eng Med Biol Soc · 2005 · 1 · 3479
24. 10.1016/j.jece.2016.03.04610.1016/j.jece.2016.03.046 ↗
25. 10.1021/jf071509h10.1021/jf071509h ↗
26. Arindam B. · 2001
27. Booth JE. · 1996
28. Siville BP. · 1999
29. Gleixner G. · Chemical Fibers Int · 2001 · 51 · 422
30. Mandot AA · Chem Fibers Int · 2016 · 66 · 190
31. Shaikh TN. · Int J Text Fashion Technol · 2013 · 3 · 9
32. Shaikh TN · Int J Emerg Sci Eng · 2015 · 3 · 11
33. Patel BH · Indian J Fibre Text Res · 2006 · 31 · 474
34. 10.1007/978-94-011-5854-1_1610.1007/978-94-011-5854-1_16 ↗