Corrosion resistance of a Nitinol ocular microstent: Implications on biocompatibility
Source abstract
Abstract
Abstract Nitinol is commonly used in medical implants due to its unique thermomechanical properties of shape memory and superelasticity. Free nickel has the potential to induce biological responses that may be a concern for permanent implants manufactured from nickel‐containing alloys. Although there are extensive reports on the effects of surface treatments on corrosion behavior in cardiovascular Nitinol implants, there is a lack of data on corrosion resistance and impact on biocompatibility for ocular implants. Therefore, the objective of this study was to determine localized corrosion and nickel elution resistance of an electropolished Nitinol‐based ocular device (Hydrus Microstent, Ivantis, Inc.) intended for patients with primary open angle glaucoma. Pitting corrosion susceptibility was characterized by potentiodynamic polarization testing per ASTM F2129. In addition, nickel ion release was quantified with immersion testing to 63 days. The results indicated high localized corrosion resistance as all samples reached polarization potentials of 800 mV without pitting initiation. Maximum nickel elution rates per device were less than approximately 1.1 ng/device/day after the first day of immersion and reduced to less than 0.1 ng/device/day after 7 days. For a patient with bilateral microstents, these nickel concentrations are ×10,000 lower than previously published tolerable intake levels for systemic toxicity. Overall, these corrosion results are in good agreement with literature values of well processed and biocompatible Nitinol devices indicating adverse systemic biological responses are not expected in vivo.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Bone Tissue Engineering Materials · Shape Memory Alloy Transformations · Ocular Disorders and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
G. RAU Inc. Scotts Valley California
References (82)
Show source-provided references
1. Study of Nitinol surface passivation using synchrotron based X‐ray photoelectron spectroscopy and absorption spectroscopy
2. Comparative in vitro biocompatibility of nickel‐titanium, pure nickel, pure titanium, and stainless steel: Genotoxicity and atomic absorption evaluation
3. 10.1016/j.athoracsur.2013.01.07910.1016/j.athoracsur.2013.01.079 ↗
4. Biocompatibility testing of NiTi screws using immunohistochemistry on sections containing metallic implants10.1002/(SICI)1097-4636(199610)32:2<243::AID-JBM14>3.0.CO;2-K ↗
5. Scanning electron microscopy analysis of corrosion of stainless steel and nitinol stents from autopsy retrievals
6. The flow of aqueous humor in the human eye
7. 10.1002/jbm.82010050510.1002/jbm.820100505 ↗
8. 10.1002/jbm.a.3072010.1002/jbm.a.30720 ↗
9. CorbettR. A.(2004). Laboratory corrosion testing of medical implants. Paper presented at Proceedings of Materials and Processes for Medical Devices Conference.
10. 10.1016/0030-4220(73)90017-010.1016/0030-4220(73)90017-0 ↗
11. 10.1007/s11665-011-9882-710.1007/s11665-011-9882-7 ↗
12. 10.1016/S1040-8428(01)00214-110.1016/S1040-8428(01)00214-1 ↗
13. 10.1016/S0921-5093(99)00294-410.1016/S0921-5093(99)00294-4 ↗
14. Surface corrosion of nitinol and stainless steel under clinical conditions
15. 10.1093/ejo/22.3.31710.1093/ejo/22.3.317 ↗
16. 10.1016/j.ajodo.2003.09.00910.1016/j.ajodo.2003.09.009 ↗
17. 10.1067/mtc.2003.12510.1067/mtc.2003.125 ↗
18. 10.1034/j.1600-0536.2000.043001035.x10.1034/j.1600-0536.2000.043001035.x ↗
19. 10.1097/IJG.000000000000001210.1097/IJG.0000000000000012 ↗
20. 10.1002/jbm.b.3169810.1002/jbm.b.31698 ↗
21. In‐vivo corrosion and local release of metallic ions from vascular stents into surrounding tissue
22. 10.1177/15266028010080030310.1177/152660280100800303 ↗
23. Biological response and biocompatibility
24. 10.1111/j.1600-0536.2008.01351.x10.1111/j.1600-0536.2008.01351.x ↗
25. 10.1007/s00403-002-0319-x10.1007/s00403-002-0319-x ↗
26. Blood flow measurements in lower limb arteries using duplex ultrasound
27. A case of nitinol allergy causing pericardial tamponade
28. 10.1016/j.jvs.2013.01.04110.1016/j.jvs.2013.01.041 ↗
29. 10.1016/S0142-9612(01)00143-010.1016/S0142-9612(01)00143-0 ↗
30. 10.1016/j.athoracsur.2009.05.04410.1016/j.athoracsur.2009.05.044 ↗
31. 10.1159/00037151110.1159/000371511 ↗
32. 10.1016/j.ijcard.2007.04.09010.1016/j.ijcard.2007.04.090 ↗
33. 10.1085/jgp.34.3.38910.1085/jgp.34.3.389 ↗
34. 10.1002/ccd.2049710.1002/ccd.20497 ↗
35. 10.1016/S0140-6736(78)92763-010.1016/S0140-6736(78)92763-0 ↗
36. 10.1007/s10856-006-0087-910.1007/s10856-006-0087-9 ↗
37. 10.1016/j.msea.2003.10.32610.1016/j.msea.2003.10.326 ↗
38. 10.1007/s40830-018-00196-010.1007/s40830-018-00196-0 ↗
39. 10.1002/jbm.b.3363010.1002/jbm.b.33630 ↗
40. 10.1016/j.actbio.2018.03.03610.1016/j.actbio.2018.03.036 ↗
41. 10.1006/taap.1998.857710.1006/taap.1998.8577 ↗
42. 10.1167/iovs.11-849610.1167/iovs.11-8496 ↗
43. 10.1038/eye.1991.3210.1038/eye.1991.32 ↗
44. 10.1093/ajcn/66.3.61610.1093/ajcn/66.3.616 ↗
45. TiNi oxidation: Kinetics and phase transformations
46. 10.1159/00012923110.1159/000129231 ↗
47. 10.1002/clc.2042710.1002/clc.20427 ↗
48. 10.1002/ccd.2215510.1002/ccd.22155 ↗
49. 10.1053/ejvs.2002.167710.1053/ejvs.2002.1677 ↗
50. 10.1111/j.1747-0803.2007.00134.x10.1111/j.1747-0803.2007.00134.x ↗
51. 10.3109/1364570000906305610.3109/13645700009063056 ↗
52. In vivo biocompatibility evaluation of nickel‐titanium shape memory metal alloy: Muscle and perineural tissue responses and encapsule membrane thickness10.1002/(SICI)1097-4636(19980905)41:3<481::AID-JBM19>3.0.CO;2-L ↗
53. Biocompatibility of nickel‐titanium shape memory metal and its corrosion behavior in human cell cultures10.1002/(SICI)1097-4636(19970615)35:4<451::AID-JBM5>3.0.CO;2-G ↗
54. 10.1016/j.ijheh.2016.10.00610.1016/j.ijheh.2016.10.006 ↗
55. 10.1016/j.actbio.2018.01.02410.1016/j.actbio.2018.01.024 ↗
56. 10.1034/j.1398-9995.2001.00086.x10.1034/j.1398-9995.2001.00086.x ↗
57. Degradation (in vitro‐in vivo corrosion)
58. 10.1111/j.1365-4632.2009.04307.x10.1111/j.1365-4632.2009.04307.x ↗
59. Evolution of the passive film on mechanically damaged Nitinol10.1002/jbm.a.32046 ↗
60. The cytotoxicity of corrosion products of nitinol stent wire on cultured smooth muscle cells10.1002/1097-4636(200011)52:2<395::AID-JBM21>3.0.CO;2-B ↗
61. 10.1016/j.amjcard.2006.12.05410.1016/j.amjcard.2006.12.054 ↗
62. 10.4236/ss.2012.3100510.4236/ss.2012.31005 ↗
63. 10.3109/1364570000906305410.3109/13645700009063054 ↗
64. 10.1007/s00330-003-2022-510.1007/s00330-003-2022-5 ↗
65. 10.1016/j.actbio.2017.08.02910.1016/j.actbio.2017.08.029 ↗
66. 10.1007/s40830-015-0028-x10.1007/s40830-015-0028-x ↗
67. Potential toxicity from nickel contamination of intravenous fluids
68. 10.1002/ajim.470014030310.1002/ajim.4700140303 ↗
69. 10.1007/978-3-642-56486-4_710.1007/978-3-642-56486-4_7 ↗
70. 10.1002/1097-4636(20000915)51:4<685::AID-JBM17>3.0.CO;2-S10.1002/1097-4636(20000915)51:4<685::AID-JBM17>3.0.CO;2-S ↗
71. Preliminary investigation of the effects of surface treatments on biological response to shape memory NiTi stents10.1002/(SICI)1097-4636(1999)48:2<165::AID-JBM11>3.0.CO;2-# ↗
72. Trepanier C. Tabrizian M. Yahia L. Bilodeau L. &Piron D.(1996). Improvement of the corrosion resistance of NiTi stents by surface treatments. Paper presented at MRS Proceedings.10.1557/PROC-459-363 ↗
73. 10.1002/(SICI)1097-4636(199824)43:4<433::AID-JBM11>3.0.CO;2-#10.1002/(SICI)1097-4636(199824)43:4<433::AID-JBM11>3.0.CO;2-# ↗
74. Trépanier C. Venugopalan R. Messer R. &Zimmerman J.2000. Effect of passivation treatments on nickel release from Nitinol. Proc Soc Biomater.
75. 10.1167/iovs.12-984110.1167/iovs.12-9841 ↗
76. 10.1002/jbm.105210.1002/jbm.1052 ↗
77. 10.1007/s11665-017-2880-710.1007/s11665-017-2880-7 ↗
78. 10.7326/0003-4819-92-5-63110.7326/0003-4819-92-5-631 ↗
79. 10.1016/S0142-9612(97)00041-010.1016/S0142-9612(97)00041-0 ↗
80. 10.1016/S0142-9612(97)00210-X10.1016/S0142-9612(97)00210-X ↗
81. Zhu L. Trepanier C. Fino J.M. Pelton A.(2004). Oxidation of Nitinol and its effect on corrosion resistance. Proceedings of ASM materials & processes for medical devices conference.
82. Zhu L. Trepanier C. Pelton A. Fino J. M.(2003). Oxidation of Nitinol and its effect on corrosion resistance. ASM Materials and Processes for Medical Devices.