Abstract
Abstract A strong, hermetic, reliable, and biocompatible ceramic‐to‐metal seal is essential for many implantable medical devices. Yttria‐stabilized tetragonal zirconia polycrystals (Y‐TZPs) and a titanium alloy Ti‐6Al‐4V were selected as the ceramic and metal components of the seal because both materials have excellent mechanical properties and favorable biocompatibility. A brazing method using titanium nickel (TiNi)‐clad braze filler material is presented to bond the components together forming a seal. Laboratory tests show that the ceramic‐to‐metal seal is hermetic, strong, and resistant to electrochemical corrosion. Twenty‐eight microstimulators utilizing the ceramic‐to‐metal seals were implanted in seven sheep to stimulate the hypoglossal nerve. When the tissue was evaluated by gross inspection at necropsy and examined histologically by a pathologist, there were no signs of local hemorrhage, infection, or hypoglossal nerve tissue damage. © 2004 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 72B: 316–321, 2005
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced materials and composites · Metal and Thin Film Mechanics · Titanium Alloys Microstructure and Properties
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (11)
Show source-provided references
1. 10.1046/j.1525-1594.2002.06938.x10.1046/j.1525-1594.2002.06938.x ↗
2. 10.1007/BF0072186410.1007/BF00721864 ↗
3. Brazing of zirconia to Ti and Ti6Al4V
4. Brazing titanium‐vapor‐coated zirconia
5. 10.1023/A:100437852052610.1023/A:1004378520526 ↗
6. Interreactions at zirconia‐Au‐Ti interfaces at high temperatures
7. Agathopoulos S · 2002 · 135
8. 10.1016/S0142-9612(02)00200-410.1016/S0142-9612(02)00200-4 ↗
9. 10.1016/S0142-9612(98)00224-510.1016/S0142-9612(98)00224-5 ↗
10. The dissociation pressures in the zirconium‐oxygen system at 1000°C
11. Park E · 1991 · 167