journal article·2022-01-26

Design and development of 3D printing assisted microwave sintering of elbow implant with biomechanical properties similar tohuman elbow

Dilpreet Singh, Bhavuk Garg, Pulak Mohan Pandey, Dinesh Kalyanasundaram

Rapid Prototyping Journal· Volume 28· Issue 2
Publisher page

Abstract

Purpose The purpose of this paper is to establish a methodology for the design and development of patient-specific elbow implant with an elastic modulus close to that of the human bone. One of the most preferred implant material is titanium alloy which is about 8 to 9 times higher in strength than that of the human bone and is the closest than other metallic biomedical materials. Design/methodology/approach The methodology begins with the design of the implant from patient-specific computed tomography information and incorporates the manufacturing of the implant via a novel rapid prototyping assisted microwave sintering process. Findings The elastic modulus and the flexural strength of the implant were observed to be comparable to that of human elbow bones. The fatigue test depicts that the implant survives the one million cycles under physiological loading conditions. Other mechanical properties such as impact energy absorption, hardness and life cycle tests were also evaluated. The implant surface promotes human cell growth and adhesion and does not cause any adverse or undesired effects i.e. no cytotoxicity. Practical implications Stress shielding, and therefore, aseptic loosening of the implant shall be avoided. In the event of any trauma post-implantation, the implant would not hurt the patient. Originality/value The present study describes a methodology for the first time to be able to obtain the strength required for the medical implant without sacrificing the fatigue life requirement.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Titanium Alloys Microstructure and Properties · Bone Tissue Engineering Materials

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Indian Institute of Technology Delhi Centre for Biomedical Engineering, , New Delhi, and Council of Scientific and Industrial Research, , Durgapur,

Central Mechanical Engineering Research Institute Centre for Biomedical Engineering, , New Delhi, and Council of Scientific and Industrial Research, , Durgapur,

All India Institute of Medical Sciences Department of Orthopedics, , New Delhi,

Indian Institute of Technology Delhi Department of Mechanical Engineering, , New Delhi,

Indian Institute of Technology Delhi Centre for Biomedical Engineering, , New Delhi, , and Department of Biomedical Engineering, , New Delhi, New Delhi,

All India Institute of Medical Sciences Centre for Biomedical Engineering, , New Delhi, , and Department of Biomedical Engineering, , New Delhi, New Delhi,

References (56)

Show source-provided references
1. Fully porous 3D printed titanium femoral stem to reduce stress-shielding following total hip arthroplasty10.1002/jor.23445 ↗
2. Behaviour of human femur bone under bending and impact loads
3. Ultrahigh-strength titanium gyroid scaffolds manufactured by selective laser melting (SLM) for bone implant applications10.1016/j.actamat.2018.08.005 ↗
4. Binderless jetting: additive manufacturing of metal parts via jetting nanoparticles
5. Direct comparison of additively manufactured porous titanium and tantalum implants towards in vivo osseointegration10.1016/j.addma.2019.04.025 ↗
6. Influence of porosity on mechanical properties and in vivo response of Ti6Al4V implants10.1016/j.actbio.2009.11.011 ↗
7. From CT scan to ceramic bone graft10.1111/j.1151-2916.2003.tb03427.x ↗
8. Reducing subsidence risk by using rapid manufactured patient-specific intervertebral disc implants10.1016/j.spinee.2012.10.003 ↗
9. Aseptic loosening of an APTIS distal radioulnar joint arthroplasty – case report10.1016/j.tcr.2019.100196 ↗
10. The effect of implant malalignment on joint loading in total elbow arthroplasty: an in vitro study10.1016/j.jse.2011.05.024 ↗
11. In vitro and in vivo performance of bioactive Ti6Al4V/TiC/HA implants fabricated by a rapid microwave sintering technique10.1016/j.msec.2014.06.015 ↗
12. Cooper · 2018
13. Metal injection moulding of titanium and titanium alloys: challenges and recent development10.1016/j.powtec.2017.06.053 ↗
14. Stress shielding of the humerus in press-fit anatomic shoulder arthroplasty: review and recommendations for evaluation10.1016/j.jse.2017.12.020 ↗
15. Mechanical behaviour of dental implants manufactured from metallic powders by μMIM10.1016/j.ctmat.2015.03.004 ↗
16. Flatfoot and normal foot a comparative analysis of the stress shielding10.1016/j.jor.2018.08.002 ↗
17. Ti based biomaterials, the ultimate choice for orthopaedic implants - A review10.1016/j.pmatsci.2008.06.004 ↗
18. Fundamentals of sintering
19. Continuous functionally graded porous titanium scaffolds manufactured by selective laser melting for bone implants10.1016/j.jmbbm.2018.01.013 ↗
20. Effect of multiple laser peening on surface integrity and microstructure of laser additive manufactured Ti6Al4V titanium alloy10.1108/RPJ-09-2018-0250 ↗
21. Results for a custom acetabular component for acetabular deficiency10.1054/arth.2002.32106 ↗
22. Strength analysis of Ti6Al4V titanium alloy produced by the use of additive manufacturing method under static load conditions10.1016/j.jmrt.2019.11.063 ↗
23. Porous metals and metal foams made from powders
24. Kondoh · 201210.5772/1922 ↗
25. Low stiffness porous Ti structures for load-bearing implants10.1016/j.actbio.2007.03.008 ↗
26. Radiographic study of carrying angle and morphometry of skeletal elements of human elbow
27. 3D-printed cellular structures for bone biomimetic implants10.1016/j.addma.2017.03.010 ↗
28. Additively manufactured 316L stainless steel with improved corrosion resistance and biological response for biomedical applications10.1016/j.addma.2019.02.005 ↗
29. Lunetto · 2020 · 101115
30. Design and manufacturing of customised femoral stems for the indian population using rapid manufacturing: a finite element approach10.3109/03091902.2011.584935 ↗
31. Diabetes mellitus and hyperglycemia and the risk of aseptic loosening in total joint arthroplasty10.1016/j.arth.2017.02.056 ↗
32. A prospective study of joint function and the quality of life of patients undergoing total joint replacement10.1111/ijn.12675 ↗
33. 3D printing in alloy design to improve biocompatibility in metallic implants10.1016/j.mattod.2020.11.021 ↗
34. Additive manufacturing (3D printing): a review of materials, methods, applications and challenges10.1016/j.compositesb.2018.02.012 ↗
35. Cold compaction and sintering of titanium and its alloys for near-net-shape or preform fabrication
37. Microstructure and mechanical behavior of porous Ti-6Al-4V processed by spherical powder sintering10.3390/ma6104868 ↗
38. Towards additive manufacturing of magnesium alloys through integration of binderless 3D printing and rapid microwave sintering10.1016/j.addma.2019.100790 ↗
39. Aseptic loosening after THA and TKA - do gender, tobacco use and BMI have an impact on implant survival time?10.1016/j.jor.2019.03.018 ↗
40. Optimization of pressure-less microwave sintering of Ti6Al4V by response surface methodology10.1080/10426914.2018.1476765 ↗
41. Microwave sintering of Ti6Al4V: optimization of processing parameters for maximal tensile strength and minimal pore size10.3390/met8121086 ↗
42. Experimental assessment of biomechanical properties in human male elbow bone subjected to bending and compression loads10.1177/2280800018793816 ↗
43. Osseointegration enhancement by Zr doping of Co-Cr-Mo implants fabricated by electron beam melting10.1016/j.addma.2015.02.002 ↗
44. Rheological behavior of PDMS silicone rubber for 3D printing of medical implants10.1016/j.addma.2018.10.004 ↗
45. Metal hypersensitivity reactions to orthopedic implants10.1007/s13555-016-0162-1 ↗
46. Surface enhancement of FDM patterns to be used in rapid investment casting for making medical implants10.1108/RPJ-07-2018-0176 ↗
47. Fatigue strength evaluation of total elbow prostheses humeral and ulnar stems
48. Design and manufacture of anatomical hip joint endoprostheses using CAD/CAM systems10.1016/S0924-0136(00)00682-8 ↗
49. TMJ concepts/techmedia custom-made TMJ total joint prosthesis: 5-year follow-up study10.1054/ijom.2002.0350 ↗
50. Processing and biocompatibility evaluation of laser processed porous titanium10.1016/j.actbio.2007.05.009 ↗
51. Microstructure, mechanical properties and superelasticity of biomedical porous NiTi alloy prepared by microwave sintering10.1016/j.msec.2014.10.053 ↗
52. Fabrication and properties of porous NiTi alloys by microwave sintering for biomedical applications10.1016/j.matlet.2014.03.088 ↗
53. Stress shielding in bone of a bone-cement interface10.1016/j.medengphy.2016.01.009 ↗
54. Enhanced fatigue behavior of powder metallurgy Ti-6Al-4V alloy by applying ultrasonic impact treatment10.1016/j.msea.2015.06.072 ↗
55. Microstructure and mechanical properties of Ti/W and Ti-6A1-4V/W composites fabricated by powder-metallurgy10.1016/S0921-5093(02)00426-4 ↗
56. Mechanical properties of the human elbow bones measured by nanoindentation and microindentation10.1115/imece2018-87406 ↗