journal article·2025-01

Management of thermal drift of bolometric infrared cameras: limits and recommendations

Samy Brazane, Olivier Riou, Fabien Delaleux, Laurent Ibos, Jean Felix Durastanti

Publisher page

Abstract

In this article, the authors propose a general framework for checking the calibration of the A325sc camera and then minimizing the temperature measurement errors due to thermal drift. Thermal drift and non-uniformity affect the measurement accuracy of infrared bolometric cameras and remain a major problem for the reproducibility and repeatability of radiance quantification. Any thermal camera is pre-calibrated at the factory to correct for thermal drift. This pre-calibration was done for specific case temperatures, and variations in these temperatures are a major source of uncertainty. To improve the accuracy of the camera measurements, it’s important to control the housing temperatures. To this end, a cold box was build up. The effectiveness of the thermal drift compensation was examined over the two ranges of the camera. In the [-20°C; 120°C] range, the thermal drift compensation is efficient up to 110°C. The range [0°C; 350°C] highlights two behaviors: for an emitting temperature within [65-225] °C, the thermal drift compensation is made difficult by the self-heating of the measurement chain due to the intensity of the source. Above 225°C, the self-heating of the optics is significant, as it becomes more absorbent. A correction of the thermosignal is suggested.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Calibration and Measurement Techniques · Infrared Target Detection Methodologies · Advanced Measurement and Metrology Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Centre de recherches en thermique, environnement et système (CERTES), IUT Sénart Fontainebleau, Université Paris Est Créteil, Créteil, France

References (14)

Show source-provided references
1. 10.1080/17686733.2015.105567510.1080/17686733.2015.1055675 ↗
2. 10.3390/s2224999710.3390/s22249997 ↗
3. 10.1117/12.54780710.1117/12.547807 ↗
4. 10.21611/qirt.2016.12610.21611/qirt.2016.126 ↗
5. 10.21611/qirt.2006.02510.21611/qirt.2006.025 ↗
6. 10.1080/17686733.2012.74606910.1080/17686733.2012.746069 ↗
7. GAVIN HP (2020). The Levenberg-Marquardt algorithm for nonlinear least squares curve-fitting problems. [Updated 2022 Nov 27
8. cited 2023 Oct 11]. Available from: https://people.duke.edu/~hpgavin/ce281/lm.pdf
9. Fluke Drywell 9100S Datasheet. [cited 2023 Oct 11]. Available from: https://docs.rs-online.com/830a/0900766b815d079c.pdf
10. Colorworks high temperature paint. [cited 2023 Oct 11]. Available from: https://docs.rs-online.com/1c36/0900766b800cb6ab.pdf
11. 10.1016/j.measurement.2007.07.00110.1016/j.measurement.2007.07.001 ↗
12. BALCON P. Spectral response of A325sc standard optic, communication personnelle. FLIR corporation; 2021.
13. GUM 1995 with minor corrections 2008. Guide to the expression of uncertainty in measurement. JCGM 100:2008. [Updated 2008 Sept; cited 2023 Oct 11]. Available from: http://www.bipm.org/utils/common/documents/jcgm/JCGM_100_2008_E.pdf
14. 10.5194/jsss-5-9-201610.5194/jsss-5-9-2016 ↗