journal article·2023-03

Development, verification and application of a versatile aerosol calibration system for online aerosol instruments

Chengfeng Liu, Cheng-Jung Huang, Kan Huang, Sheng-Hsiu Huang, Juntao Huo, Yusen Duan, Yanfen Lin

Particuology· Volume 74
Publisher page

Abstract

Traditional calibration methods mostly focus on the calibration of detection systems while the calibration from the sampling and pre-condition systems to the detection system is usually ignored. In this regard, a Primary Standard Aerosol Mass Concentration Calibration System (PAMAS) is developed for the whole-process calibration of time-resolved aerosol measurement instruments. PAMAS is composed of a particle generation chamber, an ultrasonic atomizer, a dilution system, and a syringe pump. It is designed to steadily generate standard aerosol particles of known concentrations (≤250 μg/m 3 ), chemical compositions, and stable particle size distributions. Monodispersed aerosol can be generated in the size range of hundreds of nanometers to several micrometers with a narrow size distribution. The generated particles with different compositions generated by PAMAS have been well verified by the filter-based gravimetric method, yielding accuracy and R 2 of more than 95% and 0.999 in a wide concentration range. The response time by changing the target concentration of reference particles is 1–2 min. PAMAS has been applied to various types of time-resolved aerosol measurement instruments, including particle mass concentration monitors (Beta Attenuation and Tapered Element Oscillating Microbalance), online Ion Chromatograph, and semi-continuous OCEC carbon aerosol analyzer. Very consistent results between PAMAS and calibrated instruments can be obtained if the instruments are functioning well. As for instruments with certain technical issues, PAMAS can serve as a good tool for performance evaluation and quality assurance of the instruments and the accuracy of the measurement data can be adjusted based on the calibration results.
Source: expertdata · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Atmospheric chemistry and aerosols · Air Quality Monitoring and Forecasting · Air Quality and Health Impacts

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (30)

Show source-provided references
1. A monodisperse-aerosol generation system: Design, fabrication and performance10.1016/j.partic.2017.03.005 ↗
2. Aerosol sampler operation
3. Aerosols and environmental pollution10.1007/s00114-009-0594-x ↗
4. Closure between particulate matter concentrations measured ex situ by thermal–optical analysis and in situ by the CPMA–electrometer reference mass system10.1080/02786826.2020.1788710 ↗
5. Demonstration of the CPMA-electrometer system for calibrating black carbon particulate mass instruments10.1080/02786826.2015.1010033 ↗
6. Generation and use of monodisperse aerosols
7. On the equivalence of gravimetric PM data with TEOM and beta-attenuation measurements10.1016/j.jaerosci.2004.04.004 ↗
8. Development, operation and applications of an aerosol generation, calibration and research facility special issue of aerosol science and Technology on findings from the fine particulate matter supersites Program10.1080/02786820390229516 ↗
9. Diesel bus emissions measured in a tunnel study10.1021/es030662z ↗
10. Horizontal injection spray drying aerosol generator using an ultrasonic nozzle with clean counter flow10.1016/j.jaerosci.2020.105662 ↗
11. Response characteristics of PMP compliant condensation particle counters toward various calibration aerosols10.1080/02786826.2014.1002603 ↗
12. Calibration of an aerosol composition mass spectrometer with sulfuric acid water aerosol10.1080/02786820126856 ↗
13. Development of a reliable and cost-effective weighing chamber for aerosol sample analyses10.4209/aaqr.2014.10.0260 ↗
14. Deposition of particles in turbulent pipe flows10.1016/0021-8502(94)90011-6 ↗
15. Technique for aerosol generation with controllable micrometer size distribution10.1016/j.chemosphere.2008.06.023 ↗
16. On the concentration differences between PM2.5 FEM monitors and FRM samplers10.1016/j.atmosenv.2019.117138 ↗
17. Size distributions of motor vehicle exhaust PM: A comparison between ELPI and SMPS measurements10.1080/027868200416231 ↗
18. A review of atmospheric aerosol measurements10.1016/S1352-2310(99)00455-0 ↗
19. Correction factor for continuous monitoring of Wood smoke fine particulate matter10.4209/aaqr.2010.08.0072 ↗
20. Air pollution, oxidative damage to DNA, and carcinogenesis10.1016/j.canlet.2008.02.030 ↗
21. Ultrasonic nebulization for the elemental analysis of microgram-level samples with offline aerosol mass spectrometry10.5194/amt-12-1659-2019 ↗
22. Development of an environmental chamber for evaluating the performance of low-cost air quality sensors under controlled conditions10.1016/j.atmosenv.2017.10.003 ↗
23. Multiphase chemistry at the atmosphere-biosphere interface influencing climate and public health in the anthropocene10.1021/cr500487s ↗
24. Laboratory characterization of modified tapered element oscillating microbalance samplers10.1080/10473289.2004.10471019 ↗
25. Retention and radiative forcing of black carbon in eastern Sierra Nevada snow10.5194/tc-7-365-2013 ↗
26. The CPMA-electrometer system—a suspended particle mass concentration standard10.1080/02786826.2013.801547 ↗
27. Reducing multiple charging of submicron aerosols in a corona diffusion charger10.1080/02786820701787969 ↗
28. Particle Loss Calculator – a new software tool for the assessment of the performance of aerosol inlet systems10.5194/amt-2-479-2009 ↗
29. Field performance of a semi-continuous monitor for ambient PM2.5 water-soluble inorganic ions and gases at a suburban site10.1016/j.atmosenv.2016.08.062 ↗
30. Evaluation and comparison of continuous fine particulate matter monitors for measurement of ambient aerosols10.3155/1047-3289.57.12.1499 ↗