Abstract Temporary tattoo electrodes are the most recent development in the field of cutaneous sensors. They have successfully demonstrated their performances in the monitoring of various electrophysiological signals on the skin. These epidermal electronic devices offer a conformal and imperceptible contact with the wearer while enabling good quality recordings over time. Evaluations of brain activity in clinical practice face multiple limitations, where such electrodes can provide realistic technological solutions and increase diagnostics efficiency. Here we present the performance of inkjet-printed conducting polymer tattoo electrodes in clinical electroencephalography and their compatibility with magnetoencephalography. The working mechanism of these dry sensors is investigated through the modeling of the skin/electrode impedance for better understanding of the biosignals transduction at this interface. Furthermore, a custom-made skin phantom platform demonstrates the feasibility of high-density recordings, which are essential in localizing neuropathological activities. These evaluations provide valuable input for the successful application of these ultrathin electronic tattoos sensors in multimodal brain monitoring and diagnosis.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Advanced Sensor and Energy Harvesting Materials · Neuroscience and Neural Engineering · Neural dynamics and brain function

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (67)

Show source-provided references
1. Berger, H. Über das elektrenkephalogramm des menschen. Eur. Arch. Psychiatry Clin. Neurosci. 87, 527–570 (1929).
2. Teplan, M. Fundamentals of EEG measurement. Meas. Sci. Rev. 2, 1–11 (2002).
3. Pfurtscheller, G. & Aranibar, A. Evaluation of event-related desynchronization (ERD) preceding and following voluntary self-paced movement. Electroencephalogr. Clin. Neurophysiol. 46, 138–146 (1979).10.1016/0013-4694(79)90063-4 ↗
4. Tadel, F., Baillet, S., Mosher, J. C., Pantazis, D. & Leahy, R. M. Brainstorm: a user-friendly application for MEG/EEG analysis. Comput. Intell. Neurosci. 2011, 8 (2011).10.1155/2011/879716 ↗
5. Wijdicks, E. F. M. Determining brain death in adults. Neurology 45, 1003–1011 (1995).10.1212/WNL.45.5.1003 ↗
6. Dauwels, J., Vialatte, F., Musha, T. & Cichocki, A. A comparative study of synchrony measures for the early diagnosis of Alzheimer’s disease based on EEG. NeuroImage 49, 668–693 (2010).10.1016/j.neuroimage.2009.06.056 ↗
7. Gavaret, M. et al. Source localization of scalp EEG interictal spikes in posterior cortex epilepsies investigated by HR EEG and SEEG. Epilepsia 50, 276–289 (2009).10.1111/j.1528-1167.2008.01742.x ↗
8. Müller-Putz, G. R. et al. The future in brain/neural computer interaction: Horizon 2020. p 48 (Graz: EU & Graz University of Techology, 2015).
9. Tallgren, P., Vanhatalo, S., Kaila, K. & Voipio, J. Evaluation of commercially available electrodes and gels for recording of slow EEG potentials. Clin. Neurophysiol. 116, 799–806 (2005).10.1016/j.clinph.2004.10.001 ↗
10. Ferree, T. C., Luu, P., Russell, G. S. & Tucker, D. M. Scalp electrode impedance, infection risk, and EEG data quality. Clin. Neurophysiol. 112, 536–544 (2001).10.1016/S1388-2457(00)00533-2 ↗
11. Nunez, P. L. & Srinivasan, R. Electric Fields of the Brain: The Neurophysics of EEG. (Oxford University Press, Oxford, 2006).10.1093/acprof:oso/9780195050387.001.0001 ↗
12. Lantz, G., De Peralta, R. G., Spinelli, L., Seeck, M. & Michel, C. M. Epileptic source localization with high density EEG: how many electrodes are needed? Clin. Neurophysiol. 114, 63–69 (2003).10.1016/S1388-2457(02)00337-1 ↗
13. Chi, Y. M., Jung, T.-P. & Cauwenberghs, G. Dry-contact and noncontact biopotential electrodes: methodological review. IEEE Rev. Biomed. Eng. 3, 106–119 (2010).10.1109/RBME.2010.2084078 ↗
14. Searle, A. & Kirkup, L. A direct comparison of wet, dry and insulating bioelectric recording electrodes. Physiol. Meas. 21, 271 (2000).10.1088/0967-3334/21/2/307 ↗
15. Fiedler, P. et al. in XII Mediterranean Conference on Medical and Biological Engineering and Computing 2010. 418–421 (Springer, 2010).10.1007/978-3-642-13039-7_105 ↗
16. Xing, X. et al. A high-speed SSVEP-based BCI using dry EEG electrodes. Sci. Rep. 8, 14708 (2018).
17. Stauffer, F. et al. Skin conformal polymer electrodes for clinical ECG and EEG recordings. Adv. Healthc. Mater. 7, 1700994 (2018).10.1002/adhm.201700994 ↗
18. Nawrocki, R. A. et al. Self adhesive and ultra conformable, Sub 300 nm dry thinfilm electrodes for surface monitoring of biopotentials. Adv. Funct. Mater. 28, 1803279 (2018).10.1002/adfm.201803279 ↗
19. Tian, L. et al. Large-area MRI-compatible epidermal electronic interfaces for prosthetic control and cognitive monitoring. Nat. Biomed. Eng. 3, 194 (2019).10.1038/s41551-019-0347-x ↗
20. Kabiri Ameri, S. et al. Graphene electronic tattoo sensors. ACS Nano 11, 7634–7641 (2017).10.1021/acsnano.7b02182 ↗
21. Zucca, A. et al. Tattoo conductive polymer nanosheets for skin-contact applications. Adv. Healthc. Mater. 4, 983–990 (2015).10.1002/adhm.201400761 ↗
22. Ferrari, L. M. et al. Ultraconformable temporary tattoo electrodes for electrophysiology. Adv. Sci 5, 1700771 (2018).10.1002/advs.201700771 ↗
23. da Silva, F. L. EEG and MEG: relevance to neuroscience. Neuron 80, 1112–1128 (2013).10.1016/j.neuron.2013.10.017 ↗
24. Bandodkar, A. J., Jia, W. & Wang, J. Tattoo based wearable electrochemical devices: a review. Electroanalysis 27, 562–572 (2015).10.1002/elan.201400537 ↗
25. Kim, D. H. et al. Epidermal electronics. Science 333, 838–843 (2011).10.1126/science.1206157 ↗
26. Jia, W. et al. Electrochemical tattoo biosensors for real-time noninvasive lactate monitoring in human perspiration. Anal. Chem. 85, 6553–6560 (2013).10.1021/ac401573r ↗
27. Bareket, L. et al. Temporary-tattoo for long-term high fidelity biopotential recordings. Sci. Rep. 6, 25727 (2016).10.1038/srep25727 ↗
28. Liu, Y., Pharr, M. & Salvatore, G. A. Lab-on-skin: a review of flexible and stretchable electronics for wearable health monitoring. ACS Nano 11, 9614–9635 (2017).10.1021/acsnano.7b04898 ↗
29. Chapman, B. N. Thin-film adhesion. J. Vac. Sci. Technol. 11, 106–113 (1974).10.1116/1.1318537 ↗
30. Barsotti, J. et al. Ultraconformable freestanding capacitors based on ultrathin polyvinyl formal films. Adv. Electron. Mater 4, 1800215 (2018).10.1002/aelm.201800215 ↗
31. Mehrali, M. et al. Blending electronics with the human body: a pathway toward a cybernetic future. Adv. Sci 5, 1700931 (2018).10.1002/advs.201700931 ↗
32. Sekitani, T. & Someya, T. Stretchable, large-area organic electronics. Adv. Mater. 22, 2228–2246 (2010).10.1002/adma.200904054 ↗
33. Jeong, J. W. et al. Soft materials in neuroengineering for hard problems in neuroscience. Neuron 86, 175–186 (2015).10.1016/j.neuron.2014.12.035 ↗
34. Bihar, E. et al. Fully printed electrodes on stretchable textiles for long-term electrophysiology. Adv. Mater. Technol. 2, 1600251 (2017).10.1002/admt.201600251 ↗
35. Takamatsu, S. et al. Direct patterning of organic conductors on knitted textiles for long-term electrocardiography. Sci. Rep. 5, 15003 (2015).10.1038/srep15003 ↗
36. Leleux, P. et al. Conducting polymer electrodes for electroencephalography. Adv. Healthc. Mater. 3, 490–493 (2014).10.1002/adhm.201300311 ↗
37. Arias, A. C., MacKenzie, J. D., McCulloch, I., Rivnay, J. & Salleo, A. Materials and applications for large area electronics: solution-based approaches. Chem. Rev. 110(1), 3–24 (2010).10.1021/cr900150b ↗
38. Kappenman, E. S. & Luck, S. J. The effects of electrode impedance on data quality and statistical significance in ERP recordings. Psychophysiology 47, 888–904 (2010).
39. Cole, K. S. In Cold Spring Harbor Symposia on Quantitative Biology. 110–122 (Cold Spring Harbor Laboratory Press, 2018).
40. Yamamoto, T. & Yamamoto, Y. Electrical properties of the epidermal stratum corneum. Med. Biol. Eng. 14, 151–158 (1976).10.1007/BF02478741 ↗
41. Lukacs, Z. The numerical evaluation of the distortion of EIS data due to the distribution of parameters. J. Electroanal. Chem. 432, 79–83 (1997).10.1016/S0022-0728(97)00217-9 ↗
42. Geddes, L. A. & Baker, L. E. The specific resistance of biological material—a compendium of data for the biomedical engineer and physiologist. Med. Biol. Eng. 5, 271–293 (1967).10.1007/BF02474537 ↗
43. Neuman, M. R. Biomedical Sensors. In Wiley Encyclopedia of Electrical and Electronics Engineering (ed. Webster, J. G.) (1999).10.1002/047134608X.W1408 ↗
44. Alanen, E., Nuutinen, J., Nicklén, K., Lahtinen, T. & Mönkkönen, J. Measurement of hydration in the stratum corneum with the MoistureMeter and comparison with the Corneometer. Ski. Res. Technol. 10, 32–37 (2004).10.1111/j.1600-0846.2004.00050.x ↗
45. Corcuff, P., Bertrand, C. & Leveque, J. L. Morphometry of human epidermis in vivo by real-time confocal microscopy. Arch. Dermatol. Res. 285, 475–481 (1993).10.1007/BF00376820 ↗
46. Reilly, J. P. Applied Bioelectricity: From Electrical Stimulation to Electropathology (Springer, 2012).
47. Gabriel, C. Compilation of the Dielectric Properties of Body Tissues at RF and Microwave Frequencies (King’s College London, 1996).10.21236/ADA303903 ↗
48. Björklund, S. et al. Skin membrane electrical impedance properties under the influence of a varying water gradient. Biophys. J. 104, 2639–2650 (2013).10.1016/j.bpj.2013.05.008 ↗
49. Peng, H.-L. et al. Parylene-based flexible dry electrode for bioptential recording. Sens. Actuators B 231, 1–11 (2016).10.1016/j.snb.2016.02.061 ↗
50. Hooge, F. N., Kleinpenning, T. G. M. & Vandamme, L. K. J. Experimental studies on 1/f noise. Rep. Prog. Phys. 44, 479 (1981).10.1088/0034-4885/44/5/001 ↗
51. Gilden, D. L., Thornton, T. & Mallon, M. W. 1/f noise in human cognition. Science 267, 1837 (1995).10.1126/science.7892611 ↗
52. Voytek, B. et al. Age-related changes in 1/f neural electrophysiological noise. J. Neurosci. 35, 13257–13265 (2015).10.1523/JNEUROSCI.2332-14.2015 ↗
53. Hillyard, S. A., Hink, R. F., Schwent, V. L. & Picton, T. W. Electrical signs of selective attention in the human brain. Science 182, 177–180 (1973).10.1126/science.182.4108.177 ↗
54. Rosburg, T., Boutros, N. N. & Ford, J. M. Reduced auditory evoked potential component N100 in schizophrenia—a critical review. Psychiatry Res. 161, 259–274 (2008).10.1016/j.psychres.2008.03.017 ↗
55. Puce, A. & Hämäläinen, M. A review of issues related to data acquisition and analysis in EEG/MEG studies. Brain Sci. 7, 58 (2017).10.3390/brainsci7060058 ↗
56. Hansen, P., Kringelbach, M. & Salmelin, R. MEG: An Introduction to Methods (Oxford University Press, 2010).10.1093/acprof:oso/9780195307238.001.0001 ↗
57. Badier, J. M. et al. Technical solutions for simultaneous MEG and SEEG recordings: towards routine clinical use. Physiol. Meas. 38, N118 (2017).10.1088/1361-6579/aa7655 ↗
58. Chu, C. J. High density EEG—What do we have to lose? Clin. Neurophysiol. 126, 433–434 (2015).10.1016/j.clinph.2014.07.003 ↗
59. Fender, D. H. Source localization of brain electrical activity. Handb. Electroencephalogr. Clin. Neurophysiol. 1, 355–403 (1987).
60. Lantz, G., Grouiller, F., Spinelli, L., Seeck, M. & Vulliemoz, S. Localisation of focal epileptic activity in children using high density EEG source imaging. Epileptologie 28, 84–90 (2011).
61. Srinivasan, R., Tucker, D. M. & Murias, M. Estimating the spatial Nyquist of the human EEG. Behav. Res. Methods Instrum. Comput. 30, 8–19 (1998).10.3758/BF03209412 ↗
62. Dabek, J. et al. Determination of head conductivity frequency response in vivo with optimized EIT-EEG. NeuroImage 127, 484–495 (2016).10.1016/j.neuroimage.2015.11.023 ↗
63. Gevins, A. et al. High resolution EEG: 124-channel recording, spatial deblurring and MRI integration methods. Electroencephalogr. Clin. Neurophysiol. 90, 337–358 (1994).10.1016/0013-4694(94)90050-7 ↗
64. Kandadai, M. A., Raymond, J. L. & Shaw, G. J. Comparison of electrical conductivities of various brain phantom gels: developing a ‘brain gel model’. Mater. Sci. Eng. 32, 2664–2667 (2012).10.1016/j.msec.2012.07.024 ↗
65. Jasper, H. Report of the committee on methods of clinical examination in electroencephalography. Electroencephalogr. Clin. Neurophysiol. 10, 370–375 (1958).10.1016/0013-4694(58)90053-1 ↗
66. Colombet, B., Woodman, M., Badier, J. M. & Bénar, C. G. AnyWave: a cross-platform and modular software for visualizing and processing electrophysiological signals. J. Neurosci. Methods 242, 118–126 (2015).10.1016/j.jneumeth.2015.01.017 ↗
67. Oostenveld, R., Fries, P., Maris, E. & Schoffelen, J.-M. FieldTrip: open source software for advanced analysis of MEG, EEG, and invasive electrophysiological data. Comput. Intell. Neurosci. 2011, 1687–5265 (2011).10.1155/2011/156869 ↗