A Method of Material Optimization of Cementless Stem Through Functionally Graded Material
Source abstract
Abstract
Ideally, a bone implant should be such that it exhibits an identical response to loading as real bone and is also biocompatible with existing tissue. A stiff stem, which is usually made of titanium, shields the proximal bone from mechanical loading (stress shielding). On the other hand, decreasing the stem stiffness increases the proximal interface shear stress and the risk of proximal interface failure. Therefore the purpose of this study is to solve these conflicting requirements in order to have more uniform interface shear stress distribution and less stress shielding through the concept of functionally graded material (FGM). FGM is a kind of advanced composite materials, which changes its composition and structure gradually over one or two directions of its volume, resulting in corresponding changes in the properties of the material. This study is divided into two parts; in the first part, the finite element analysis and optimization technique are used to design the stem as one-dimensional FGM, while in the second part, the stem is designed as two-dimensional functionally graded material. The aim of both designs is to overcome the above mentioned problems. In the case of part one (one-dimensional FGM), the gradation of elastic modulus is changed along the vertical direction (model 1) and along the horizontal direction (model 2), in order to find the optimal gradation direction. It is found that the optimal design is to change the elastic modulus gradually from 110 GPa (Hydroxyapatite) at the top of the stem to 1GPa (Collagen) at the bottom (model 1). This optimal gradation decreases stress shielding by 83%, while reduces the maximum interface shear stress by 32% compared to homogenous titanium stem. However, in the second part (two-dimensional FGM, model 3) the materials of optimal design are found to be hydroxyapatite, Bioglass, and collagen. This design leads to the same stress shielding reduction as in model 1, while at the same time, the maximum interface shear stress is reduced by 45% and 63% compared to the optimal one-dimensional FGM design and homogenous titanium stem, respectively.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Bone Tissue Engineering Materials · Orthopaedic implants and arthroplasty · Titanium Alloys Microstructure and Properties
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (28)
Show source-provided references
1. M. Akay · Journal of Biomedical Materials Research · 1996 · 31 · 16710.1002/(SICI)1097-4636(199606)31:2<167::AID-JBM3>3.0.CO;2-L ↗
2. W. Cao · Ceramics International · 1996 · 22 · 49310.1016/0272-8842(95)00126-3 ↗
3. Gatti, A.M., Hench, L.L., Morari, E., Thompson, I., and Tanza, D., (1998). Bioceramics. in R.Z. LeGeros, J.P. LeGeros (Eds.) World Scientific, Singapore, Vol. 11. pp. 349–352.
4. S. Gross · Journal of Biomechanics · 2001 · 34 · 99510.1016/S0021-9290(01)00072-0 ↗
5. H.S. Hedia · Bio-Medical Materials and Engineering · 2000 · 11 · 293
6. H.S. Hedia · Bio-Medical Materials and Engineering · 2000 · 10 · 73
7. H.S. Hedia · Bio-Medical Materials and Engineering · 1997 · 7 · 8310.3233/BME-1997-7201 ↗
8. H.S. Hedia · Bio-Medical Materials and Engineering · 1996a · 6 · 19910.3233/BME-1996-6307 ↗
9. H.S. Hedia · Medical Engineering & Physics · 1996b · 18 · 64710.1016/S1350-4533(96)00025-2 ↗
10. H.S. Hedia · Bio-Medical Materials and Engineering · 2004 · 14 · 133
11. R. Huiskes · Journal of Biomechanics · 1989 · 22 · 79310.1016/0021-9290(89)90063-8 ↗
12. Itoh, S., Kawauchi, T., Shinomiya, K., Koyama, Y., et al. (2000). Osteoconductive activity of a novel biomaterial, Hydroxyapatite/Collagen composite, 46th Annual Meeting, March 12–15. Orthopaedic Research Society, Orlando, Florida.
13. S. Itoh · Hydroxyapatite/collagen composite, Biomaterials · 2002 · 23 · 391910.1016/S0142-9612(02)00126-6 ↗
14. Jeffers, J.R.T. and Taylor, M., (2003). Mesh Considerations for Adaptive Finite Element Analyses of Cement Failure in Total Hip Replacement. Sonesta Beach Resort in Key Biscayne, Florida, Summer Bioengineering Conference, June 25–29.
15. J.R. Jones · Material Science and Technology · 2001 · 17 · 89110.1179/026708301101510762 ↗
16. H. Katoozian · Medical Engineering & Physics · 2001 · 23 · 50310.1016/S1350-4533(01)00079-0 ↗
17. T. Kokubo · Journal of Biomaterials · 2003 · 24 · 216110.1016/S0142-9612(03)00044-9 ↗
18. Kuiper J.H., Huiskes R., (1992). Numerical optimization of hip-prosthetic stem material, Recent Advances in Computer Methods in Biomechanics & Biomedical Engineering. J. Mddleton, G.N. Pande, and K.R. Williams, pp. 76–84.
19. J.H. Kuiper · Journal of Biomechanical Engineering · 1997 · 119 · 16610.1115/1.2796076 ↗
20. R. Miyao · The Journal of the Japanese Society for Dental Materials and Devices · 2001 · 20 · 344
21. M. Nematt-Allah · International Journal of Solids and Structures · 2003 · 40 · 733910.1016/j.ijsolstr.2003.08.017 ↗
22. W. Pompe · Ceramic Transactions · 2001 · 114 · 65
23. R. Shirandami · Journal of Biomedical Engineering · 1990 · 12 · 1910.1016/0141-5425(90)90109-Z ↗
24. J.A. Simoes · Composite Science and Technology · 2000 · 60 · 55910.1016/S0266-3538(99)00155-4 ↗
25. Thompson, I., Hench, L.N., (2000). Comprehensive composite materials Vol. 6: Design and applications, 6.39: Medical applications of composites, Elsevier Science Ltd., pp. 727–753.
26. F. Watari · Materials Science Forum · 2003 · 423–425 · 32110.4028/www.scientific.net/MSF.423-425.321 ↗
27. H. Weinans · Journal of Biomechanics · 2000 · 33 · 80910.1016/S0021-9290(00)00036-1 ↗
28. L.J. Zhang · Materials Letters · 2004 · 58 · 71910.1016/j.matlet.2003.07.009 ↗