journal article·2009-05

Hip stem fatigue test prediction

Heidi-Lynn Ploeg, Maja Bürgi, Urs P. Wyss

International Journal of Fatigue· Volume 31· Issue 5
Publisher page

Abstract

Abstract The accuracy of fatigue test prediction methods for the standard fatigue testing of hip stems was evaluated against the experimental results of static and fatigue tests. Axial unnotched strain-controlled material fatigue tests provided the required cyclic material properties. Finite element analysis of the hip stems predicted a maximum tensile stress to within 3–7% of strain gauge measurements. The four methods investigated accurately predicted hip stem fatigue strength at 5 million cycles (−1% to −9% errors). The strain–life methods successfully predicted fatigue life (factors 1/7.0–9.2 of the test) at high and low stress amplitudes of 352 and 315 MPa, respectively. The classical stress–life method was only accurate (factor 1/1.9) for the low stress level. The current study has demonstrated that fatigue test prediction methods can be applied with confidence to support standard fatigue testing of hip stems. Further studies can expand the understanding of these methods and their clinical relevance by investigating effects due to variable amplitude loading and environment.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Hip disorders and treatments · Total Knee Arthroplasty Outcomes

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References

Reference metadata was not supplied for this record.