journal article·2018-08

Metallurgical aspects of failure in a broken femoral HIP prosthesis

M. Kiani Khouzani, A. Bahrami, A. Eslami

Publisher page

Abstract

Abstract In this research failure of femoral hip prosthesis, made of AISI 316 L stainless steel, in a 69 years old patient with a weight of 70 kg was investigated. Replacement of the damaged cartilage and bone of the hip joint with prosthesis component through Total HIP Arthroplasty (THA) is a widely used surgery. Patients with this type of implants are anticipated to have a normal daily life with capability of performing routine physical activities for a period of 15 years. However, in investigated case, the failure had occurred in
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Hydrogen embrittlement and corrosion behaviors in metals · Advanced materials and composites

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (22)

Show source-provided references
1. HIP replacement, landmark surgery in modern history10.1016/j.maturitas.2013.04.011 ↗
2. Fatigue failure analysis of a specific total hip prosthesis stem design10.1016/j.ijfatigue.2007.11.005 ↗
3. Bopdhb
4. Data Bridge Market Research
5. Texture development and microstructure evolution in metastable austenitic steel processed by accumulative roll bonding and subsequent annealing10.1007/s10853-014-8254-8 ↗
6. Material failure of an AISI 316L stainless steel hip prosthesis
7. Mechanical and corrosion properties of forged hip stems made of high-nitrogen stainless steel
8. Failure analysis of a femoral hip stem made of stainless steel after short time of exposure
9. Fracture of cementless femoral stems at the Mid-Stem Junction in Modular Revision hip arthroplasty systems10.2106/JBJS.I.01589 ↗
10. Case study and failure analysis of a total hip stem fracture
11. Case study failure of a high nitrogen stainless steel femoral stem10.1016/j.engfailanal.2011.11.010 ↗
12. Analysis of microstructural aspects of a hip stem failure made of the REX 734 stainless steel10.1016/j.engfailanal.2016.11.012 ↗
13. Failure analysis of a total hip prosthesis implanted in active patient10.1016/j.jmbbm.2010.06.004 ↗
14. Premature failure of a femoral hip replacement10.1016/j.matpr.2016.03.058 ↗
15. Campbell · 2012
16. Case study and analysis of a fatigue failure in a THA stem10.1016/j.engfailanal.2012.10.011 ↗
17. Narayan · 2012 · vol. 23
18. Recent development in stainless steels. J10.1016/j.mser.2009.03.001 ↗
19. Sigma phase-induced failure of AISI 310 stainless steel radiant tubes10.1016/j.engfailanal.2017.08.027 ↗
20. A case study on failure of AISI 347H stabilized stainless steel pipe in a petrochemical plant10.1016/j.csefa.2017.07.001 ↗
21. Fatigue fracture in dual modular revision total hip arthroplasty stems failure analysis and computed tomography diagnostics in two cases10.1016/j.arth.2013.09.008 ↗
22. Graef · 2007