journal article·2013-02-01

A Study on Wear Testing of Orthopedic Implant Materials in Simulated Body Fluid

Erdem Atar

Materials Testing· Volume 55· Issue 2
Publisher page

Abstract

Abstract In this study, the sliding wear behaviour of the materials utilized in manufacturing of load-bearing orthopedic implants (316 L, Ti6Al4V, and CoCrMo nickel alloys) was examined by a reciprocating wear tester under ceramic-on-metal configuration in a simulated body fluid (SBF). During the tests the generated wear debris were transferred into the SBF and increased its optical density. In accordance with its poor wear resistance, Ti6Al4V alloy provided a large amount of wear debris transfer into the SBF along with the heavy material attachment to the contact surface of the counterface. Dissolution of wear debris led to an increase of the metal ion concentration in the SBFs. Even though few wear debris were generated, trace elements were detected in the SBFs of the 316 L and CoCrMo alloy, besides the main elements.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Titanium Alloys Microstructure and Properties · Advanced materials and composites

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Gebze, Kocaeli, Turkey

References (40)

Show source-provided references
1. 2003
2. 2004
3. 2010 · 2 · 40
4. Electrochemical corrosion of metal implants · 2005 · 184 · 1219
5. The problem of corrosion in orthopedic implant materials · 1999 · 9 · 1
6. Characterization of different materials for corrosion resistance under simulated body fluid conditions · 2002 · 49 · 73
7. 1998 · 80A · 268
8. Metallic biomaterials of knee and hip – A review · 2010 · 24 · 68
9. The combined role of wear particles · 2008 · 5 · 1263
10. Aseptic loosening · 2006 · 77 · 177
11. 2003 · 1
12. Inflammatory bone loss in joint replacements: The mechanisms · 2010 · 26 · 1
13. Inflammatory bone loss in joint replacements: The key mediators · 2010 · 26 · 1
14. Evaluation of the biological response of wear debris · 2004 · 4 · 239S
15. How useful is SBF in predicting in vivo bone bioactivity · 2006 · 27 · 2907
16. Enhanced wear resistance of titanium surfaces by a new thermal oxidation treatment · 2000 · 238 · 131
17. A brief review of test methodologies for surface-engineered biomedical implant alloys · 2005 · 198 · 483
18. 483
19. Metal ion release from metal implants · 2004 · 24 · 745
20. An analysis of the in vivo deterioration of CoCrMo implants through wear and corrosion · 2007 · 221 · 291
21. 1997 · 79B · 316
22. Comparision of metal release from various metallic biomaterials in vitro · 2005 · 26 · 11
23. 2010 · 180
24. Metal release from stainless steel · 2008 · 50 · 3429
25. Nickel-free austenitic stainless steels for medical applications · 2010 · 11 · 1
26. Comparative evaluation of metal ions release from titanium and Ti6Al7Nb into bio-fluids · 2009 · 6 · 7
27. Wear and corrosion behaviour of Ti13Nb13Zr and Ti6Al4V alloys in simulated physiological solution · 2011 · 53 · 796
28. Titanium alloy mini-implants for orthodontic anchorage: immediate loading and metal ion release · 2007 · 3 · 331
29. 2009 · 67 · 182
30. 2004
31. 2008 · 31
32. Corrosion degradation and prevention by surface modification of biometallic materials · 2007 · 18 · 725
33. Titanium alloy mini-implants for orthodontic anchorage: Immediate loading and metal ion release · 2007 · 3 · 331
34. Ti based biomaterials · 2009 · 54 · 397
35. Surface Modification for Bioimplants: The Role of Laser Surface Engineering · 2005 · 20 · 5
36. Functional Coatings of Films for Hard-Tissue applications · 2010 · 3 · 3994
37. Surface modification of titanium alloys for combined improvements in corrosion and wear resistance · 1998 · 107 · 125
38. Surface modification of titanium · 2004 · R47 · 49
39. Effect of thermal oxidation on corrosion and corrosion-wear behaviour of a Ti6Al4V alloy · 2004 · 25 · 3325
40. 2006 · 26 · 1367