journal article·2020Open access

Titanium-Niobium (Ti-xNb) Alloys with High Nb Amounts for Applications in Biomaterials

Bruno Leandro Pereira, Carlos Maurício Lepienski, Viviane Seba, Guibert Hobold, Paulo Soares, Bor Shin Chee, Pedro Akira Bazaglia Kuroda, Erico Saito Szameitat, Leonardo Luís dos Santos, Carlos Roberto Grandini, Michael Nugent

Materials Research· Volume 23· Issue 6

Abstract

The present study produced binary titanium-niobium alloys Ti-xNb with high niobium percentage (x = 50, 80 e 90 wt.%) aiming to apply it as a base material in osseous implant devices. The produced TiNb alloys presented a cubic crystalline phase, lower corrosion rate than titanium, lower elastic modulus values, and superior cellular viability than titanium and niobium. The Ti50Nb alloy among all analyzed metallic substrates showed the best values of elastic modulus, corrosion properties, wettability, and cellular viability. Thus, this work suggests a Ti/Nb ratio close to 1 (Ti50Nb) displays optimum characteristics to apply in orthopedic devices.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Titanium Alloys Microstructure and Properties · Bone Tissue Engineering Materials · Orthopaedic implants and arthroplasty

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Universidade Federal do Paraná, Brasil; Athlone Institute of Technology, Ireland

Universidade Federal do Paraná, Brasil; Universidade Federal do Paraná, Brasil; Universidade Tecnológica Federal do Paraná, Brasil

Athlone Institute of Technology, Ireland

Pontifícia Universidade Católica do Paraná, Brasil

Universidade Tecnológica Federal do Paraná, Brasil; Colorado State University, United States of America

Universidade Estadual Paulista, Brasil

Universidade Federal do Paraná, Brasil

References (24)

Show source-provided references
1. Davis JR · 2003
2. Metallic implant biomaterials10.1016/j.mser.2014.10.001 ↗
3. Musa A · 2003
4. Recent metallic materials for biomedical applications10.1007/s11661-002-0109-2 ↗
5. Problem of stress shielding and improvement to the hip implant designs: a review10.3923/jms.2007.460.467 ↗
6. Biocompatibility and osteogenesis of refractory metal implants, titanium, hafnium, niobium, tantalum and rhenium10.1016/S0142-9612(00)00275-1 ↗
7. Ti based biomaterials, the ultimate choice for orthopaedic implants: a review10.1016/j.pmatsci.2008.06.004 ↗
8. Lütjering G · 2007
9. Ab-initio and experimental study of phase stability of Ti-Nb alloys10.1016/j.jallcom.2016.11.231 ↗
10. Mechanical and biological properties of Ti-(0-25 wt%)Nb alloys for biomedical implants application10.1093/rb/rbz042 ↗
11. Characterization, corrosion behavior, cellular response and in vivo bone tissue compatibility of titanium-niobium alloy with low Young’s modulus10.1016/j.msec.2015.10.062 ↗
12. Beta Ti-45Nb and Ti-50Nb alloys produced by powder metallurgy for aerospace application10.4028/www.scientific.net/MSF.660-661.405 ↗
13. Influence of the substitutional solute on the mechanical properties of Ti-Nb binary alloys for biomedical use10.1590/S1516-14392012005000040 ↗
14. Electrochemical and bioactive characteristics of the porous surface formed on Ti-xNb alloys via plasma electrolytic oxidation10.1016/j.surfcoat.2019.125027 ↗
15. Factors influencing the elastic moduli, reversible strains and hysteresis loops in martensitic Ti-Nb alloys10.1016/j.msec.2014.12.048 ↗
16. Biocompatibility of Ti-alloys for long-term implantation10.1016/j.jmbbm.2012.11.014 ↗
17. Bica VHF · 2019 · 1
18. 2020
19. 2015
20. Corrosion behaviour of polished and sandblasted titanium alloys in PBS solution
21. Ponte H. A. · 2003
22. Attraction between hydrated hydrophilic surfaces10.1016/j.cplett.2014.07.046 ↗
23. Effect of wettability and surface functional groups on protein adsorption and cell adhesion using well-defined mixed self-assembled monolayers10.1016/j.biomaterials.2007.03.013 ↗
24. Osteoblast adhesion on biomaterials10.1016/S0142-9612(99)00242-2 ↗