journal article·2025-04Open access

The influence of microstructure on the corrosion behavior of platinum used for cochlea implant electrodes

Tatiana Blank, Sabine Behrens, Elisabeth Franziska Lange, Christian Klose, Demircan Canadinc, Martin Durisin, Thomas Lenarz, Hans Jürgen Maier

Corrosion Science· Volume 246

Abstract

Cochlear implants (CIs) have revolutionized the treatment of profound hearing loss, yet their long-term durability remains a challenge due to potential corrosion-induced degradation of the platinum electrodes of the CI. This study examines the corrosion behavior of platinum wires under accelerated test conditions, and compares the results with corrosion patterns observed in explanted CI electrodes from patients after almost 1 year and 26 years. Accelerated corrosion tests with frequencies from 1 to 5 kHz were validated as predictive tools for long-term performance, as the surface morphologies of stimulated platinum wires exhibit notable similarities to those of explanted electrodes. The as-drawn wires analyzed in this study exhibit up to 50 % more corrosion attack than annealed ones. Additionally, the impact of grain orientation on corrosion is examined by testing platinum single crystals with different electrochemical methods. For this purpose, single crystals with the orientation [111], [110] and [100] were analyzed. The research highlights the importance of annealing platinum and minimizing plastic deformation during electrode assembly to enhance electrode durability. Corrosion was found to primarily occur at grain boundaries, with grain orientation showing minimal influence. These findings suggest that engineering grain boundary characteristics could significantly improve the corrosion resistance of polycrystalline electrodes, offering a promising pathway to enhance the longevity of CIs. • Experiments and explants showed similar corrosion, proving their usefulness in predicting long-term in-service behavior. • The as-drawn platinum wires showed more pronounced corrosion attack in all accelerated corrosion tests. • Corrosion attack was most pronounced at the grain boundaries. • Annealed platinum and minimize deformation during assembly to enhance the electrode's longevity.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Electrodeposition and Electroless Coatings · Corrosion Behavior and Inhibition · Metal and Thin Film Mechanics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (37)

Show source-provided references
1. Lenarz · 1998
2. Cochlear implant – state of the art10.1055/s-0043-101812 ↗
3. A. Huber, Cochlea-Implantation 21 (2021) 474–478. https://doi.org/08765.
4. Platinum in biomedical applications
5. Platinum corrosion products from electrode contacts of human cochlear implants induce cell death in cell culture models10.1371/journal.pone.0196649 ↗
6. Cellular immunologic responses to cochlear implantation in the human10.1016/j.heares.2014.09.007 ↗
7. Foreign body response to silicone in cochlear implant electrodes in the human10.1097/MAO.0000000000001454 ↗
8. Cochlear implants: causes, effects and mitigation strategies for the foreign body response and inflammation10.1016/j.heares.2022.108536 ↗
9. Predicting corrosion delamination failure in active implantable medical devices: analytical model and validation strategy10.3390/bioengineering9010010 ↗
10. Impedance changes and fibrous tissue growth after cochlear implantation are correlated and can be reduced using a dexamethasone eluting electrode10.1371/journal.pone.0147552 ↗
11. Platinum dissolution and tissue response following long-term electrical stimulation at high charge densities10.1088/1741-2552/abe5ba ↗
12. Chronic intracochlear electrical stimulation at high charge densities: reducing platinum dissolution10.1088/1741-2552/abb7a6 ↗
13. The role of grain boundaries in the corrosion process of fe surface: insights from ReaxFF molecular dynamic simulations10.3390/met12050876 ↗
14. Effect of grain boundary and crystallographic orientation on the stress corrosion behavior of an Al-Zn-Mg alloy10.1007/s11665-019-04050-x ↗
15. Inhibition of galvanic corrosion between crystallographic orientations in low alloy steel by grain ultra-refinement
16. Influence of grain orientation on the corrosion behavior of rolled AZ31 magnesium alloy10.1007/s40195-015-0337-2 ↗
17. Cochlear implants: system design, integration, and evaluation10.1109/RBME.2008.2008250 ↗
18. Microstructure and mechanical behavior of annealed MP35N alloy wire10.1016/j.msea.2015.03.103 ↗
19. Radixin modulates the function of outer hair cell stereocilia10.1038/s42003-020-01506-y ↗
20. Electrical stimulation with Pt electrodes: trace analysis for dissolved platinum and other dissolved electrochemical products10.1159/000124611 ↗
21. Coulometric detection of irreversible electrochemical reactions occurring at Pt microelectrodes used for neural stimulation10.1021/ac103037u ↗
22. Corrosion of platinum electrodes in phosphate buffered saline solution10.1002/maco.19970480506 ↗
23. Study on the microstructure, mechanical properties and corrosion behavior of Mg-Zn-Ca alloy wire for biomaterial application10.1007/s11665-018-3278-x ↗
24. Microstructural impact on intergranular corrosion and the mechanical properties of industrial drawn 6056 aluminum wires10.1016/j.matdes.2015.05.079 ↗
25. Correlation of microstructural and mechanical properties of K-doped tungsten fibers used as reinforcement of tungsten matrix for high temperature applications10.1016/j.ijrmhm.2018.12.007 ↗
26. Deformation and recrystallised texture evolution and the followed mechanical and electrical properties of drawn and annealed copper wires10.4028/www.scientific.net/JERA.31.20 ↗
27. Effect of drawing and annealing on the microstructure and mechanical properties of 304 austenitic stainless steel wire10.1088/2053-1591/ac44d6 ↗
28. Solubility of platinum in aqueous solutions at 25°C and pHs 4 to 10 under oxidizing conditions10.1016/S0016-7037(01)00752-9 ↗
29. Sulfuric acid solutions of Pt(OH)4(H2O)2: a platinum speciation survey and hydrated Pt(IV) oxide formation for practical use10.1021/acs.inorgchem.2c01134 ↗
30. MED-EL, The MED-EL Blog (20 April 2016).
31. Stress corrosion cracking resistant nanostructured Al-Mg alloy with low angle grain boundaries10.1016/j.actamat.2021.117607 ↗
32. P. Pedeferri, Intergranular and Selective Corrosion, Corrosion Science and Engineering (2018) 297–311. https://doi.org/10.1007/978-3-319-97625-9_15.10.1007/978-3-319-97625-9_15 ↗
33. Accelerated anode and cathode reaction due to direct electron uptake and consumption by manganese dioxide and titanium dioxide composite cathode in degradation of iron composite10.1016/j.jcis.2022.11.055 ↗
34. Correlative study of Mott–Schottky analysis and corrosion behaviour of plasma electrolytic oxidation treated magnesium alloy: Effect of Cl⎺ concentrations10.1016/j.corsci.2024.112386 ↗
35. Correlating corrosion to surface grain orientations of polycrystalline aluminum alloy by scanning electrochemical cell microscopy10.1021/acsami.2c12813 ↗
36. Crystallographic anisotropy of corrosion rate and surface faceting of polycrystalline 90Cu-10Ni in acidic NaCl solution10.1016/j.matdes.2022.110429 ↗
37. Effect of crystal orientations and precipitates on corrosion behavior of Al-Cu-Li single crystals10.1016/S1003-6326(22)66065-5 ↗