Corrosion and metal release rates of Ni substrates under a porous gold coating
Source abstract
Abstract
Abstract Noble metal coatings intended to protect devices from corrosion may have small porosity that allows corrosive fluid to attack the substrate alloy. A technique combining electrochemical tests and chemical analysis is described that measures low levels of porosity and low corrosion product release rates from the substrate. The technique has been applied to an electrodeposited gold coating on a nickel substrate. The results yielded a porosity of 1.8 × 10−3, defined as the area of substrate exposed to corrosion/area of the coating. Galvanic coupling between gold and nickel was found to increase the rate of nickel corrosion by about a factor of 10. A simplified model showed that if the substrate alloy does not passivate, the corrosion product release rate is expected to increase with exposure time.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Electrodeposition and Electroless Coatings · Corrosion Behavior and Inhibition · Anodic Oxide Films and Nanostructures
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (6)
Show source-provided references
1. Corrosion behavior of titanium nitride coated Ni–Ti shape-memory surgical alloy10.1016/S0142-9612(00)00368-9 ↗
2. Electrochemical corrosion and metal ion release from Co–Cr–Mo prosthesis with titanium plasma spray coating10.1016/j.biomaterials.2005.01.004 ↗
3. Corrosion resistance, mechanical properties, corrosion fatigue strength and cytocompatibility of new Ti alloys without Al and V10.1016/S0142-9612(97)00235-4 ↗
4. Comparison of metal release rates from various metallic biomaterials in vitro10.1016/j.biomaterials.2004.02.005 ↗
5. Nickel release after implantation of the Amplatzer occluder10.1067/mhj.2003.7 ↗
6. Ion release from orthodontic wires in artificial saliva with various acidities10.1016/S0142-9612(03)00188-1 ↗