journal article·2020-10-31Open access

Effects of Various Polishing Techniques on the Surface Characteristics of the Ti-6Al-4V Alloy and on Bacterial Adhesion

Yu-Ting Jhong, Chih-Yeh Chao, Wei-Chun Hung, Je-Kang Du

Coatings· Volume 10· Issue 11

Abstract

Ti-6Al-4V, although widely used in dental materials, causes peri-implant inflammation due to the long-term accumulation of bacteria around the implant, resulting in bone loss and eventual failure of the implant. This study aims to overcome the problem of dental implant infection by analyzing the influence of Ti-6Al-4V surface characteristics on the quantity of accumulated bacteria. Ti-6Al-4V specimens, each with different surface roughness are produced by mechanical, chemical, and electrolytic polishing. The surface roughness, surface contact angle, surface oxygen content, and surface structure were measured via atomic force microscopy (AFM), laser scanning confocal microscopy (LSCM), drop shape analysis (using sessile drop), X-ray photoelectron spectroscopy (XPS), and X-ray diffraction (XRD). The micro and macro surface roughness are 10.33–120.05 nm and 0.68–2.34 μm, respectively. The surface X direction and Y direction contact angle are 21.38°–96.44° and 18.37°–92.72°, respectively. The surface oxygen content is 47.36–59.89 at.%. The number of colonies and the optical density (OD) are 7.87 × 106–17.73 × 106 CFU/mL and 0.189–0.245, respectively. The bacterial inhibition were the most effective under the electrolytic polishing of Ti-6Al-4V. The electrolytic polishing of Ti-6Al-4V exhibited the best surface characteristics: the surface roughness of 10 nm, surface contact angle of 92°, and surface oxygen content of 54 at.%, respectively. This provides the best surface treatment of Ti-6Al-4V in dental implants.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Bone Tissue Engineering Materials · Dental materials and restorations · Dental Implant Techniques and Outcomes

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Ph.D. Program, School of Dentistry, College of Dental Medicine, Kaohsiung Medical University, Kaohsiung 80708, Taiwan

Department of Mechanical Engineering, National Pingtung University of Science and Technology, Pingtung 91201, Taiwan

Department of Microbiology and Immunology, College of Medicine, Kaohsiung Medical University, Kaohsiung 80708, Taiwan

School of Dentistry, College of Dental Medicine, Kaohsiung Medical University, Kaohsiung 80708, Taiwan

Department of Dentistry, Kaohsiung Medical University Hospital, Kaohsiung 80708, Taiwan

Funding

Kaohsiung Medical University Chung-Ho Memorial Hospital · Ministry of Science and Technology of the People's Republic of China

References (50)

Show source-provided references
1. Recent metallic materials for biomedical applications10.1007/s11661-002-0109-2 ↗
2. Aging response of the young’s modulus and mechanical properties of Ti-29Nb-13Ta-4.6Zr for biomedical applications10.1007/s11661-003-0230-x ↗
3. Failure causes, timing, and cluster behavior: An 8-year study of dental implants10.1097/ID.0b013e3181777906 ↗
4. Periodontal disease as a risk for dental implant failure over time: A long-term historical cohort study10.1111/j.1600-051X.2011.01745.x ↗
5. Characteristics of early versus late implant failure: A retrospective study10.1016/j.joms.2009.07.050 ↗
6. Treatment of infections associated with surgical implants10.1056/NEJMra035415 ↗
7. Influence of surface modifications to titanium on antibacterial activity in vitro10.1016/S0142-9612(00)00392-6 ↗
8. Effect of polishing method on surface roughness and bacterial adhesion of zirconia-porcelain veneer10.1016/j.ceramint.2016.11.036 ↗
9. Corrosion behaviour of titanium in the presence of Streptococcus mutans10.1016/j.jdent.2013.03.008 ↗
10. Antibiotic resistance in human peri-implantitis microbiota10.1111/clr.12160 ↗
11. In vitro assessment of stainless steel orthodontic brackets coated with titanium oxide mixed Ag for anti-adherent and antibacterial properties against Streptococcus mutans and Porphyromonas gingivalis10.1016/j.micpath.2017.09.052 ↗
12. Antibacterial activity, corrosion resistance and wear behavior of spark plasma sintered Ta-5Cu alloy for biomedical applications10.1016/j.jmbbm.2017.06.022 ↗
13. Influence of titanium and zirconia modified surfaces for rapid healing on adhesion and biofilm formation of Staphylococcus epidermidis10.1016/j.archoralbio.2020.104824 ↗
14. Streptococcus mutans attachment on a cast titanium surface10.1590/S1516-14392009000100003 ↗
15. Impact of nano-topography on bacterial attachment10.1002/biot.200700244 ↗
16. The influence of abutment surface roughness on plaque accumulation and peri-implant mucositis10.1034/j.1600-0501.1996.070302.x ↗
17. Bacterial adhesion to glass and metal-oxide surfaces10.1016/j.colsurfb.2004.05.006 ↗
18. Properties of the stainless steel substrate, influencing the adhesion of thermo-resistant streptococci10.1016/S0260-8774(99)00157-0 ↗
19. Morphology, chemical composition, and electrochemical characteristics of colored titanium passive layers10.1021/la001694s ↗
20. Inactivation of Staphylococcus aureus and Escherichia coli under various light sources on photocatalytic titanium dioxide thin film10.1016/j.surfcoat.2008.09.036 ↗
21. A functionalized TiO2/Mg2TiO4 nano-layer on biodegradable magnesium implant enables superior bone-implant integration and bacterial disinfection10.1016/j.biomaterials.2019.119372 ↗
22. Microstructure, antibacterial properties and wear resistance of plasma Cu–Ni surface modified titanium10.1016/j.surfcoat.2013.06.012 ↗
23. Assessment of cytotoxicity and antibacterial effects of silver nanoparticle-doped titanium alloy surfaces10.1016/j.dental.2019.06.003 ↗
24. Antibacterial effect of metallic glasses10.1007/s11434-012-4997-2 ↗
25. Antithrombogenic investigation and biological behavior of cultured human umbilical vein endothelial cells on Ti-O film10.1007/s11431-004-5108-7 ↗
26. Towards long-lasting antibacterial stainless steel surfaces by combining double glow plasma silvering with active screen plasma nitriding10.1016/j.actbio.2010.08.009 ↗
27. Bactericidal behavior of Cu-containing stainless steel surfaces10.1016/j.apsusc.2012.06.074 ↗
28. Study on antimicrobial activity of chitosan with different molecular weights10.1016/j.carbpol.2003.07.009 ↗
29. Interface shear strength of titanium implants with a sandblasted and acid-etched surface: A biomechanical study in the maxilla of miniature pigs10.1002/(SICI)1097-4636(199905)45:2<75::AID-JBM1>3.0.CO;2-P ↗
30. Direct assessment of profilometric roughness variability from typical implant surface types
31. The effect of HF-HNO3 chemical polishing on the surface roughness and fatigue life of laser powder bed fusion produced Ti6Al4V10.1016/j.mtcomm.2020.101396 ↗
32. Bone response to surface modified titanium implants: Studies on electropolished implants with different oxide thicknesses and morphology10.1016/0142-9612(94)90092-2 ↗
33. Implant infections: Adhesion, biofilm formation and immune evasion10.1038/s41579-018-0019-y ↗
34. Electropolishing and electropolishing-related allowances for powder bed selectively laser-melted Ti-6Al-4V alloy components10.1016/j.jmatprotec.2016.11.014 ↗
35. Electrochemical polishing of additively manufactured Ti–6Al–4V alloy10.1007/s12540-019-00556-0 ↗
36. Characterization of surface finish of electropolished stainless steel AISI 316L with varying electrolyte concentrations10.1016/j.proeng.2013.08.255 ↗
37. The effects of stainless steel finish on Salmonella Typhimurium attachment, biofilm formation and sensitivity to chlorine10.1016/j.fm.2013.02.005 ↗
38. Influence of surface roughness on streptococcal adhesion forces to composite resins10.1016/j.dental.2011.03.017 ↗
39. The influence of substratum topography on bacterial adhesion to polymethyl methacrylate10.1023/A:1008874326324 ↗
40. Metal surface characteristics dictate bacterial adhesion capacity10.1016/j.ijadhadh.2016.01.008 ↗
41. The influence of surface roughness and surface-free energy on supra- and subgingival plaque formation in man10.1111/j.1600-051X.1995.tb01765.x ↗
42. Microbial changes in biofilms on composite resins with different surface roughness: An in vitro study with a multispecies biofilm model10.1016/j.prosdent.2019.08.009 ↗
43. Assessment of hydrophobicity and roughness of stainless steel adhered by an isolate of Bacillus cereus from a dairy plant10.1590/S1517-83822010000400017 ↗
44. Direction dependent dynamic wetting of semi-hemispherical end micro-groove textured Ti-6Al-4V surface10.1016/j.surfcoat.2018.09.037 ↗
45. Effect of hydrophobicity on in vitro streptococcal adhesion to dental alloys10.1023/A:1008913915399 ↗
46. The influence of nano-scale surface roughness on bacterial adhesion to ultrafine-grained titanium10.1016/j.biomaterials.2010.01.071 ↗
47. Surface optical second harmonic generation from RutileTiO2(110) in air10.1143/JJAP.36.7250 ↗
48. Surface characterization of implant materials cp Ti, Ti-6Al-7Nb and Ti-6Al-4V with different pretreatments10.1023/A:1008840026907 ↗
49. Anomalous enhancement of corrosion resistance and antibacterial property of commercially pure Titanium (CP-Ti) with nanoscale rutile titania film10.1016/j.corsci.2020.108678 ↗
50. The electropolishing of titanium and its alloys10.1016/S0026-0576(04)82510-8 ↗